Measurement Guideline 1
44
Measurement Guideline 1
Connector Considerations
CAUTION: The RF present at any Test Set input connector must not exceed the specified
level or permanent instrument damage may result. If necessary, use an external
attenuator. If overpower occurs, disconnect the Transmitter, then cycle Test Set
power OFF/ON to reset the protection circuitry.
RF IN/OUT
The RF present at the Test Set RF IN/OUT connector must not exceed
60W continuous (or 100 Watts for 10 sec/minute) or permanent
instrument damage may result.
ANT IN
The RF present at the Test Set ANT IN connector must not exceed 200
mW or permanent instrument damage may result.