186 Chapter 2
Performance Verification Tests
23. Frequency Response, Agilent E4402B and E4403B
and 2.0 GHz. Record this number in Table 2-48 as the Minimum
Response for Band 0B.
26.For Band 0B in Table 2-48, subtract the Minimum Response value
from the Maximum Response value and record the result in the
Peak-to-Peak Response column.
27.Transfer the values entered in Table 2-48 to the test record in this
guide.
28.The frequency response test for analyzers with Option UKB is now
complete.
Table 2-47 Frequency Response Results, dc Coupled
Band Maximum Response Minimum Response Peak-to-Peak Response
dB
Test
Record
Entry
dB
Test
Record
Entry
dB
Test
Record
Entry
01) 2) 3)
0A 4) 5) 6)
0B 7) 8) 9)
Table 2-48 Frequency Response Results, ac Coupled
Band
Maximum Response Minimum Response Peak-to-Peak Response
dB
Test
Record
Entry
dB
Test
Record
Entry
dB
Test
Record
Entry
0 10) 11) 12)
0A 13) 14) 15)
0B 16) 17) 18)