Temperature
Operating 10 °C to +40 °C (50 °F to +104 °F)
Non-operating –40 °C to +65 °C (–40 °F to +158 °F)
Humidity
Operating Up to 90% humidity (non-condensing) at +40 °C (+104 °F)
Non-operating Up to 95% relative humidity at +65 °C (+149 °F)
Altitude
Operating Up to 4,600 meters (15,000 ft)
Non-operating Up to 15,300 meters (50,000 ft)
Vibration
Operating Random vibration 5 to 500 Hz, 10 minutes per axis, 0.21 g (rms)
Non-operating Random vibration 5 to 500 Hz, 10 minutes per axis, 0.3 g (rms); Resonant search, 5 to 500 Hz
swept sine, 1 octave/min sweep rate, 0.5 g, 5 minute resonant dwell at 4 resonances/axis
Power requirements
Voltage 90 to 132 or 198 to 264 Vac, 48 to 66 Hz
Power (including modules) 604 VA; 391 W
Weight
Mainframe without modules 15.5 kg (34 lb)
Typical module 1.2 kg (2.6 lb)
Mainframe dimensions (excluding handle)
Without front connectors and rear feet 215.1 mm H x 425.5 mm W x 566 mm D (8.47 in x 16.75 in x 22.2 in)
With front connectors and rear feet 215.1 mm H x 425.5 mm W x 629 mm D (8.47 in x 16.75 in x 24.8 in)
Specifications
Mainframe specifications
HORIZONTAL SYSTEM (time base) PATTERN LOCK
Scale factor (full scale is ten divisions)
Minimum 2 ps/div (with 86107A: 500 fs/div)
Maximum 1 s/div 250 ns/div
Delay
1
Minimum 24 ns 40.1 ns
Maximum 1000 screen diameters or 10 s, 1000 screen diameters or 25.401 µs,
whichever is smaller whichever is smaller
Time interval accuracy
2
1 ps + 1.0% of ∆ time reading
3
8 ps + 0.1% of ∆ time reading
Time interval accuracy – jitter mode operation
4
1 ps
Time interval accuracy – with 86107A < 200 fs
precision timebase
Time interval resolution ≤ (screen diameter)/(record length) or 62.5 fs,
whichever is larger
Display units Bits or time (TDR mode–meters)
VERTICAL SYSTEM (channels)
Number of channels 4 (simultaneous acquisition)
Vertical resolution 14 bit A/D converter (up to 15 bits with averaging)
Full resolution channel scales Adjusts in a 1-2-5-10 sequence for coarse adjustment or fine adjustment resolution
from the front panel knob
Adjustments Scale, offset, activate filter, sampler bandwidth, attenuation factor, transducer conversion factors
Record length 16 to 4096 samples – increments of 1
10
Specifications describe warranted performance over the temperature range of +10 °C to +40 °C (unless otherwise noted). The specifications are
applicable for the temperature after the instrument is turned on for one (1) hour, and while self-calibration is valid. Many performance parameters are
enhanced through frequent, simple user calibrations. Characteristics provide useful, non-warranted information about the functions and
performance of the instrument. Characteristics are printed in italic typeface.
Factory Calibration Cycle -For optimum performance, the instrument should have a complete verification of specifications once every twelve (12) months.
General specifications
This instrument meets Agilent Technologies’ environmental specifications (section 750) for class B-1 products with exception as described for temperature and
condensation. Contact your local field engineer for complete details. Product specifications and descriptions in this document subject to change without notice.
1
Time offset relative to the front panel trigger input on the instrument mainframe.
2
Dual marker measurement performed at a temperature within ±5 °C of horizontal calibration temperature.
3
Delay settings: ∆ time is in the range (26 + N*4 ns) ±1.9 ns, where N = 0, 1, 2, ... 17.
4
Characteristic performance. Test configuration: PRBS of length 2
7
– 1 bits, Data and Clock 10 Gb/s.