OUTPut Commands SCPI Commands Reference
Agilent 81133A/81134A Pulse Generator Programming Guide, March 2007 39
OUTPut Commands
The following table shows the Agilent 81133A/81134A Pulse
Generator OUTPut commands.
Command Parameter Description
:OUTPut
[0|1|2]:DIVider[?] Numeric | MIN |
MAX
Set/read channel frequency
divider
[0]:SOURce[?] PERiodic |
BITStream
Set/read trigger source mode
[0|1|2][:STATe][?] ON | OFF | 1 | 0
Set/read channel outputs on
and off
[1|2]:NEG[?] ON | OFF | 1 | 0
Set/read negative channel
output on and off
[1|2]:POS[?] ON | OFF | 1 | 0
Set/read positive channel out-
put on and off
:CENTral[?] ON | OFF | 1 | 0
Set/read central output set-
tings