Agilent Technologies 54503A Betriebsanweisung Seite 101

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 716
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 100
Chapter 2 101
Making Measurements
Example 8: External Millimeter Mixers (Unpreselected)
Figure 2-45 Select the band of interest.
Save the average conversion-loss value
4. Table lists default conversion-loss values that are stored in the
analyzer for each frequency band. These values approximate the
values for the Agilent 11970 series mixers. Other conversion-loss
values may be entered into the spectrum analyzer in two ways. The
first method lets you to save the average conversion-loss value for
the entire band using
AVERAGE CNV LOSS. To activate this function:
a. Press
AUX CTRL, EXTERNAL MIXER, AMPTD CORRECT, AVERAGE
CNV LOSS.
b. Enter the appropriate average conversion-loss value. On Agilent
11970 Series harmonic mixers, these values are charted on the
mixer. The U-band mixer used for this example had an average
conversion loss of 23.5 dB, as shown in Figure 2-46.
Seitenansicht 100
1 2 ... 96 97 98 99 100 101 102 103 104 105 106 ... 715 716

Kommentare zu diesen Handbüchern

Keine Kommentare