Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test
for 0 m Capacitance Measurement Accuracy Test (DC Bias: ON) (Page 9-41) to the following
information.
Standard
Test Signal
Frequency
Parameter
Measured
Test Limits
Test Result
1
Measurement
Uncertainty
1 μF 1 k Cp ±0.0011 μF μF ± 0.000066 μF
D ±0.0011 ± 0.000033
Change 8
Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test
for 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: LONG) (Page 9-42) to
the following information.
Standard
Test Signal
Frequency
Parameter
Measured
Test Limits
Test Result
1
Measurement
Uncertainty
100 mohm 100 Hz R ±0.52 mohm mohm ± 0.030 mohm
100 mohm 1 kHz R ±0.48 mohm mohm ± 0.030 mohm
Change 9
Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test
for 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: MED) (Page 9-42) to the
following information.
Standard
Test Signal
Frequency
Parameter
Measured
Test Limits
Test Result
1
Measurement
Uncertainty
100 mohm 100 Hz R ±0.62 mohm mohm ± 0.033 mohm
Change 10
Change the Measurement Uncertainty value of 0 m Impedance Measurement Accuracy Test
for 0 m Resistance Measurement Accuracy Test (Meas. Time Mode: SHORT) (Page 9-42) to
the following information.
Standard
Test Signal
Frequency
Parameter
Measured
Test Limits
Test Result
1
Measurement
Uncertainty
100 mohm 100 Hz R ±1.10 mohm mohm ± 0.13 mohm