114 Agilent E6474A User’s Guide
6 Get Assistance, if You Need It
If your system is equipped with a navigator, perform the tests in
Table 5 on page 116 next. If you experience problems in setting
up the tests, refer to “Troubleshooting your E6474A system" on
page 102.
Test 3 - Receiver testing by measuring the noise floor
Preparing the hardware
1 Connect the system as shown in Figure 54.
2 Turn on the Agilent receiver.
NOTE
Use this procedure only if you do not have a signal generator and there is
no known channel in the system. Connect a 50-ohm load (customer
supplied) to the input of the receiver.
Figure 54 System Configuration for Measuring the Noise Floor