Seite 1 - Waveform Generator

Agilent TechnologiesAgilent U2761A USB Modular Function/Arbitrary Waveform GeneratorService Guide

Seite 2

x U2761A Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Seite 3 - Safety Symbols

U2761A Service Guide xiContents1 Characteristics and Specifications 1Product Characteristics 2Product Specifications and Characteristics 42 G

Seite 4 - General Safety Information

xii U2761A Service Guide Calibration Procedure 30Internal Timebase Adjustment 31Output Impedance Adjustment 32DC Offset (High Impedance) Adju

Seite 5 - Environment Conditions

U2761A Service Guide xiiiList of FiguresFigure 3-1 U2761A to frequency counter connection 18Figure 3-3 U2761A to DMM connection 20Figure 3

Seite 6 - Regulatory Markings

xiv U2761A Service Guide List of TablesTab l e 3- 1 Recommended test equipment 16Tab l e 3- 2 Configuration for timebase verification

Seite 7 - 2002/96/EC

1Agilent U2761A USB Modular Function/Arbitrary Waveform GeneratorService GuideAgilent Technologies1

Seite 8 - In This Guide…

1 Characteristics and Specifications 2 U2761A Service GuideProduct CharacteristicsREMOTE INTERF

Seite 9

Characteristics and Specifications 1 U2761A Service Guide 3[1] Requires a direct USB connection to the PC so the appropriate driver can be installed i

Seite 10

1 Characteristics and Specifications 4 U2761A Service GuideProduct Specifications and Character

Seite 11 - Contents

Characteristics and Specifications 1 U2761A Service Guide 5Ramp, TriangleFrequency range 1 μHz to 200 kHz (1 μHz resolution)Linearity < 0.2% of pea

Seite 12

ii U2761A Service GuideNotices© Agilent Technologies, Inc., 2008–2013No part of this manual may be reproduced in any form or by any means (including e

Seite 13 - List of Figures

1 Characteristics and Specifications 6 U2761A Service GuideCOMMON CHARACTERISTICSAmplitudeRange

Seite 14 - List of Tables

Characteristics and Specifications 1 U2761A Service Guide 7MODULATIONAMCarrier waveforms Sine, Square, Ramp, ArbitrarySource InternalInternal modulati

Seite 15 - Agilent Technologies

1 Characteristics and Specifications 8 U2761A Service Guide1 Add 1/10th of output amplitude an

Seite 16 - Product Characteristics

9Agilent TechnologiesAgilent U2761A USB Modular Function/Arbitrary Waveform GeneratorService Guide2Getting StartedIntroduction 10Self-Test 10Agi

Seite 17 - U2761A Service Guide 3

2 Calibration 10 U2761A Service GuideIntroductionSelf-TestA brief power-on self-test occurs aut

Seite 18 - WAVEFORM CHARACTERISTICS

Calibration 2 U2761A Service Guide 112 Click Start on the Self-Test form.3 Click OK to continue when the message box appears.4 Wait for a few m

Seite 19 - Arbitrary

2 Calibration 12 U2761A Service GuideAgilent Technologies Calibration ServicesWhen your U2761A

Seite 20 - Range +360 ° to –360 °

Calibration 2 U2761A Service Guide 13Obtaining Repair Service (Worldwide)To obtain service for your U2761A (in-warranty, under service contract,

Seite 21

2 Calibration 14 U2761A Service GuideRepackaging for ShipmentIf the U2761A is to be shipped to

Seite 22 - Trigger output

15Agilent TechnologiesAgilent U2761A USB Modular Function/Arbitrary Waveform GeneratorService Guide3CalibrationIntroduction 16Recommended Test Equi

Seite 23 - Getting Started

U2761A Service Guide iiiSafety SymbolsThe following symbols on the instrument and in the documentation indicate precautions which must be taken to mai

Seite 24 - Introduction

3 Calibration 16 U2761A Service GuideIntroductionIf the U2761A fails any of the tests or if an

Seite 25

Calibration 3 U2761A Service Guide 17Test ConsiderationFor optimum performance, all procedures should comply with the following recommendations.• Ensu

Seite 26 - Extended Service Contracts

3 Calibration 18 U2761A Service GuidePerformance Verification ProceduresIt is recommended to c

Seite 27

Calibration 3 U2761A Service Guide 192 Set the U2761A to the output shown in the table below and measure the output frequency. Make sure that the U276

Seite 28 - Cleaning

3 Calibration 20 U2761A Service Guide2 Use the DMM to make the resistance measurement at the o

Seite 29 - Calibration

Calibration 3 U2761A Service Guide 212 Set the U2761A based on the following.• Load impedance: High-Z• Function: DC3 Set the U2761A to each output lis

Seite 30

3 Calibration 22 U2761A Service Guide3 Set the U2761A to each output listed in the table below

Seite 31 - Test Consideration

Calibration 3 U2761A Service Guide 23Amplitude Flatness (–20 dB Range) VerificationThis procedure verifies the high frequency AC amplitude flatness ab

Seite 32

3 Calibration 24 U2761A Service Guide2 Set the U2761A based on the following.• Load impedance:

Seite 33 - Output Impedance Verification

Calibration 3 U2761A Service Guide 25Table 3 - 7 Configuration for AC amplitude flatness (–20 dB range) verification U2761A MeasurementFunction Amplit

Seite 34

iv U2761A Service GuideGeneral Safety InformationThe following general safety precautions must be observed during all phases of operation of this inst

Seite 35 - Nominal Error

3 Calibration 26 U2761A Service GuideAmplitude Flatness (0 dB Range) VerificationThis procedur

Seite 36

Calibration 3 U2761A Service Guide 27Table 3 - 8 Configuration for AC amplitude flatness (0 dB range) verification U2761A MeasurementFunction Amplitud

Seite 37 - + 20 dB

3 Calibration 28 U2761A Service GuideAmplitude Flatness (+20 dB Range) VerificationThis proced

Seite 38 - to the Offset

Calibration 3 U2761A Service Guide 29Table 3 - 9 Configuration for AC amplitude flatness (+20 dB range) verification U2761A MeasurementFunction Amplit

Seite 39

3 Calibration 30 U2761A Service GuideCalibration ProcedureCalibration is necessary when the on

Seite 40 - to Offset

Calibration 3 U2761A Service Guide 31Please note that the calibration constant will only be stored after the user completes the subsequent calibration

Seite 41

3 Calibration 32 U2761A Service Guide5 Proceed to output impedance adjustment or exit the cali

Seite 42

Calibration 3 U2761A Service Guide 331 Calibration constant is stored after completing this setup.5 Proceed to DC offset adjustment or exit the calib

Seite 43

3 Calibration 34 U2761A Service Guide4 Proceed to AC amplitude (High Impedance) adjustment or

Seite 44 - Calibration Procedure

Calibration 3 U2761A Service Guide 354 Fill in the table below using the measurement results in Table 3- 13.Table 3 - 14 Sine wave flatness worksheet5

Seite 45 - Internal Timebase Adjustment

U2761A Service Guide vEnvironment ConditionsThis instrument is designed for indoor use and in the area with low condensation. The table below shows th

Seite 46 - Output Impedance Adjustment

3 Calibration 36 U2761A Service Guide7 Use the DMM to perform the AC voltage measurement at th

Seite 47

Calibration 3 U2761A Service Guide 37Amplitude Flatness (–20 dB Range) AdjustmentThis procedure calibrates the AC amplitude flatness for high frequenc

Seite 48

3 Calibration 38 U2761A Service Guide1 Calibration constant is stored after completing this s

Seite 49

Calibration 3 U2761A Service Guide 39Amplitude Flatness (0 dB Range) AdjustmentThis procedure calibrates the AC amplitude flatness for high frequency

Seite 50

3 Calibration 40 U2761A Service Guide1 Calibration constant is stored after completing this s

Seite 51

Calibration 3 U2761A Service Guide 41Amplitude Flatness (+20 dB Range) AdjustmentThis procedure calibrates the AC amplitude flatness for high frequenc

Seite 52

3 Calibration 42 U2761A Service Guide1 Calibration constant is stored after completing this s

Seite 53

43Agilent TechnologiesAgilent U2761A USB Modular Function/Arbitrary Waveform GeneratorService Guide4 DisassemblyGeneral Disassembly 44Disassembly I

Seite 54

4 Disassembly 44 U2761A Service GuideGeneral DisassemblyThis chapter provides the step-by-step

Seite 55

Disassembly 4 U2761A Service Guide 45Remove the two indicated screws from the front metal piece.Gently pull the front metal piece out, which is attach

Seite 56

vi U2761A Service GuideRegulatory MarkingsThe CE mark is a registered trademarkof the European Community.This CEmark shows that the product complieswi

Seite 57 - Disassembly

4 Disassembly 46 U2761A Service GuideReassembly InstructionsThe reassembly process is simply t

Seite 58 - General Disassembly

Disassembly 4 U2761A Service Guide 473 You can search for the replacement parts by entering a specific replacement part number or by instruments.• To

Seite 59

4 Disassembly 48 U2761A Service GuideTHIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK.

Seite 60 - Replacement Parts

49Agilent U2761A USB Modular Function/Arbitrary Waveform GeneratorService GuideAgilent Technologies5

Seite 61

5 Troubleshooting 50 U2761A Service GuideTroubleshootingThis section offers suggestions for s

Seite 62 - 4 Disassembly

U2761A Service Guide 51IndexAaccuracy, 5, 12, 16, 18, 19, 20adjustment+20 dB range flatness, 400 dB range flatness, 38-20 db range AC flatness, 36AC a

Seite 63 - Troubleshooting

52 U2761A Service GuideIndexPphase noise, 3PM 6carrier waveform, 6phase deviation, 6power consumption, 2power meter, 16, 22, 23, 25, 27, 36, 38, 40pow

Seite 64

© Agilent Technologies, Inc. 2008–2013Seventh Edition, August 2, 2013U2761-90010www.agilent.comAgilent TechnologiesContact usTo obtain service, warran

Seite 65 - U2761A Service Guide 51

U2761A Service Guide viiWaste Electrical and Electronic Equipment (WEEE) Directive 2002/96/ECThis instrument complies with the WEEE Directive (2002/96

Seite 66 - 52 U2761A Service Guide

viii U2761A Service GuideIn This Guide…1 Characteristics and SpecificationsThe characteristics and specifications of the U2761A are listed in this cha

Seite 67

U2761A Service Guide ixDeclaration of Conformity (DoC)The Declaration of Conformity (DoC) for this instrument is available on the Web site. You can se

Kommentare zu diesen Handbüchern

Keine Kommentare