Troubleshooting
–299 to –200: Execution Error Messages
–299 to –200:
Execution Error Messages
An error number in the range [–299 to –200] indicates that an error has
been detected during instrument execution.
Execution Error Message Descriptions
–221 Settings conflict; parameter currently
disabled
This parameter is grayed out (unavailable) in the
current context. Check the individual parameter
help/documentation for more information.
–223 Too much data; <description of the type of
data exceeded>
Indicates that a legal program data element of block,
expression or string type was received that contained
more data than the device could handle due to memory
or related device-specific requirements.
–230 Data corrupt or stale.
Possibly invalid data. A new measurement was started
but not completed.