About this Manual
We’ve added this manual to the Agilent website in an effort to help you support
your product. This manual is the best copy we could find; it may be incomplete
or contain dated information. If we find a more recent copy in the future, we will
add it to the Agilent website.
Support for Your Product
Agilent no longer sells or supports this product. Our service centers may be able
to perform calibration if no repair parts are needed, but no other support from
Agilent is available. You will find any other available product information on the
Agilent Test & Measurement website,
www.tm.agilent.com.
HP References in this Manual
This manual may contain references to HP or Hewlett-Packard. Please note that
Hewlett-Packard's former test and measurement, semiconductor products and
chemical analysis businesses are now part of Agilent Technologies. We have
made no changes to this manual copy. In other documentation, to reduce
potential confusion, the only change to product numbers and names has been in
the company name prefix: where a product number/name was HP XXXX the
current name/number is now Agilent XXXX. For example, model number
HP8648A is now model number Agilent 8648A.
Seitenansicht 0
1 2 3 4 5 6 ... 300 301

Inhaltsverzeichnis

Seite 1

About this Manual We’ve added this manual to the Agilent website in an effort to help you support your product. This manual is the best copy we co

Seite 2 - Oscilloscope

Testing Threshold Accuracy 7–12To set up and connect the equipment 7–13To verify threshold accuracy 7–14To check the output of the CALIBRATOR 7–15

Seite 3 - The HP 54645A/D at a Glance

Displaying Analog Channels After the AcquisitionThe two scope channels are always acquired, so you can turn them on and look at them even if they were

Seite 4

Pan and ZoomThe ability to pan (move horizontally) and zoom (expand or compresshorizontally) an acquired waveform is important because of the addition

Seite 5

In peak detect display mode, the useful limit is more apparent from thedisplay. As we zoom in on peak detected data, we eventually see theresolution o

Seite 6 - Contents

With vectors off, the resolution limit appears as shown in the following figure.Resolution Limit in Peak Detect with Vectors OffRecall that in peak de

Seite 7

Recovering the waveform on the screenIn the HP 54645A/D, pressing Run/Stop freezes the screen. Wheneither of the following occur, there will be seve

Seite 8

Changing the time reference position1Press Main/Delayed .2Choose the time reference position from the Time Reference softkeys:•To set the time referen

Seite 9

The time reference position sets the initial position of the trigger event withinacquisition memory and on the display, with delay set to 0. The delay

Seite 10 - 8 Replacement Parts

At 5 µs/div, we are sampling at 200 MSa/s (scope), and 400 MSa/s (digital).When no delay value is set, and the time reference is set to Center, thetri

Seite 11 - Glossary

When we pan the display (move it horizontally) back and set the delay to15 µs, the trigger point is displayed on the screen. The part of the signal we

Seite 12 - Getting Started

Setting a negative delay lets us look at the part of the signal to the left of thetrigger point. Suppose we set the time reference to Right, with a sw

Seite 13

To remove the power supply 8–8To remove the keyboard 8–9To remove the handle 8–9Replacing the Option 005 board 8–10Replaceable Parts 8–11To order a re

Seite 14 - Figure 1-1 Figure 1-2

If we set the time reference to center, then pan the display back to set thedelay to 0 µs, the trigger point is displayed at center screen, and the re

Seite 15 - Preparing the Oscilloscope

Post Acquisition ProcessingIn addition to changing display parameters after the acquisition, all of themeasurements and math functions can be done aft

Seite 16 - Check package contents

Run/Stop/Single/Autostore/Erase OperationDuring an acquisition, the oscilloscope examines the input voltage ateach input probe. •For analog channels o

Seite 17 - Figure 1-3

Acquiring DataThe HP 54645A/D operates like an analog scope, but it can do much more.Spending a few minutes to learn some of this capability will take

Seite 18 - Figure 1-4

Memory Depth/Record LengthRun/Stop versus SingleThe HP 54645A/D both have a Single key and a Run/Stop key. When youpress Run, the trigger processing o

Seite 19 - Power Cords

Given that this is the case, we now have two options:•We could acquire a full 500,000 points for every acquisition, resulting insome samples being “of

Seite 20 - Connect the cables

To run and stop an acquisition•To begin an acquisition, press the Run/Stop key.The instrument begins acquiring data while searching for a trigger con

Seite 21 - Power-on the oscilloscope

To capture a single eventTo capture a single event, you need some previous knowledge of the signal toset up the trigger level and slope. For example,

Seite 22

4Press Single .Pressing Single arms the trigger circuit. When the trigger conditionsare met, data appears on the display, representing the data points

Seite 23

To use AutostoreWith Autostore or infinite persistence, the scope updates the display withnew acquisitions, but does not erase the results of previous

Seite 25

To Autostore Multiple Single EventsTo Autostore single events, you first need to configure the instrument to dosingle Autostores.1Press Autostore to

Seite 26 - Front-Panel Overview

To erase the waveform display•Press Erase .Acquisition memory and the current display are immediately erased. If theinstrument is in Run or Autostore

Seite 27

Using Digital Channels to Probe CircuitsYou may encounter problems when using the HP 54645D that are related toprobing. These problems fall into two c

Seite 28

High-Frequency Probe Equivalent CircuitThe impedance plots for the two models are shown in these figures. Bycomparing the two plots, you can see that

Seite 29

Probe GroundingA probe ground is the low-impedance path for current to return to the sourcefrom the probe. Increased length in this path will, at high

Seite 30

In addition to the common mode voltage—ground bounce—longer groundreturns also degrade the pulse fidelity of the probe system. Risetime isincreased, a

Seite 31 - HP 54645A/D Front Panels

Saving and Recalling the ConfigurationWith the HP 54645A/D Oscilloscopes, you can save traces and theoscilloscope setup. Traces are stored in trace me

Seite 32 - Figure 2-2

Trace Memory Softkeys on HP 54645D with HP 54659B Measurement/Storage Module InstalledSetup memoryThe HP 54645A/D has 10 separate setup memories that

Seite 33 - HP 54645A Front-Panel Areas

For example, suppose that you use the HP 54645A/D in field testing of asystem where ten separate tests are used. You can predefine the testconfigurati

Seite 34 - HP 54645D Front-Panel Areas

To save, recall, and clear tracesTo save traces, two trace memories are available. The following exerciseshows you how to store and recall waveforms f

Seite 35 - Table 2-3

Getting StartedWhen you use the HP 54645A/D Oscilloscopes to help test andtroubleshoot your systems, you will do the following:• Prepare the oscillosc

Seite 36

3Select a trace memory by pressing the Trace softkey.4Press the Save to Mem<N> softkey.The current oscilloscope screen and setup are copied to t

Seite 37

To save or recall front-panel setupsTen memories are available to store front-panel setups. Saving front-panelsetups can save you time in situations w

Seite 38

Table 4-1HP 54645A Factory-Preset Default Configuration SettingsConfiguration Item SettingCursors Cursors off; time readout is selected; all cursors a

Seite 39

Table 4-2HP 54645D Factory-Preset Default Configuration SettingsConfiguration Item SettingCursors Cursors off; time readout is selected; all cursors a

Seite 40

Configuring the Mixed-Signal OscilloscopeThis section describes how to configure the following items for theHP 54645D Mixed-Signal Oscilloscope, inclu

Seite 41 - Front-Panel Operation

Logic Threshold SetupThe threshold voltage setting is used by the input comparators to determinewhether an input signal is a logic low or logic high.

Seite 42

Using Labels on the Mixed-Signal OscilloscopeThe HP 54645D Mixed-Signal Oscilloscope allows you to define andassign labels to each input channel. Or,

Seite 43

To assign a predefined label to a channel1Press Label/Threshold .2Press the Define Labels softkey.The following figure shows the label maker. A label

Seite 44 - , then press

If you defined a new label, it is added to the non-volatile label list and will besaved with the oscilloscope configuration.HP 54645D Label Definition

Seite 45

Label List ManagementThe label list contains 75 of the most recently used labels. The list does not saveduplicate labels, nor does it save multiple la

Seite 46

The HP 54645A/D high-speed display can be used to isolateinfrequently changing signals. You can then use the characteristics ofthese signals to help r

Seite 47

4To enter characters into the new label:aPress the Position softkey until the cursor position in the entry fieldcorresponds to the point where you wan

Seite 48 - , press

To initialize the label list1Press Label/Threshold .2Press the Initialize Label List softkey.A message appears, warning you that this operation will o

Seite 49

5Making Measurements withHP 54645A/D Oscilloscopes

Seite 50

Making Measurements with HP 54645A/DOscilloscopesBy now you are familiar with the VERTICAL/ANALOG, HORIZONTAL, andTRIGGER groups of the front-panel ke

Seite 51 - D0-D15

Capturing DataIn the HP 54645A Oscilloscope, an 8-bit A/D converter converts theinput waveform into an evenly-spaced series of voltage readings.These

Seite 52

To use delayed sweepYou can use the delayed sweep window to locate and horizontally expandpart of the main sweep for a more detailed (high-resolution)

Seite 53

The screen divides in half. The top half displays the main sweep, and thebottom half displays an expanded portion of the main sweep. This expandedpo

Seite 54

To view asynchronous noise on a signalThe following exercise shows how to use the oscilloscope to viewasynchronous noise on a signal that is not synch

Seite 55 - To make cursor measurements

2Press Autostore .Notice that STORE is displayed in the status line.3Set the trigger mode to normal, then adjust the trigger level into thenoise regio

Seite 56

To reduce the random noise on a signalIf the signal you are applying to the oscilloscope is noisy, you can set up theoscilloscope to reduce the noise

Seite 57

Preparing the OscilloscopeTo prepare your HP 54645A Oscilloscope or HP 54645D Mixed-SignalOscilloscope for use, you need to do the following tasks. Af

Seite 58 - To modify the graticule

Low-frequency reject (LF reject) adds a high-pass filter with the 3-dB pointat 50 kHz. Use LF reject to remove low-frequency signals, such as power l

Seite 59 - To print the display

3Use averaging to reduce noise on the displayed waveform.•Press Display , then press Average.Notice that Av appears in the status line.•Toggle the # A

Seite 60

To capture glitches or narrow pulses with peak detectA glitch is a rapid change in the waveform that is usually narrow as comparedto the waveform. Pea

Seite 61 - Glitch Trigger

3Characterize the glitch with delayed sweep.To characterize the glitch with delayed sweep, follow these steps:•Press Main/Delayed . Press the Delayed

Seite 62 - Mode/Coupling

With the Measurement/Storage module installed, two functions define theseoperations that create mathematically altered waveforms (not pixel math):•Fun

Seite 63

To use the XY display modeThe XY display mode converts the oscilloscope from a volts-versus-timedisplay to a volts-versus-volts display. You can use

Seite 64 - To use holdoff

Signal Centered on the Display4Press Cursors .5Set the Y2 cursor to the top of the signal, and set Y1 to the bottom ofthe signal.Note the ∆Y value at

Seite 65

Cursors Set to Center of Signal6Move the Y1 and Y2 cursors to the center of the signal.Again, note the ∆Y value.7Calculate the phase difference using

Seite 66 - To use external triggering

Signals are 90° Out of PhaseSignals are In PhaseFigure 5-12Figure 5-13Making Measurements with HP 54645A/D OscilloscopesTo use the XY display mode5-17

Seite 67 - To trigger on an edge

To analyze video waveformsThe TV sync separator in the oscilloscope has an internal clamp circuit. Thisremoves the need for external clamping when yo

Seite 68 - To use glitch triggering

Check package contents•Verify that you received the following items and any optionalaccessories in the HP 54645A/D packaging.If anything is missing, c

Seite 69 - To use TV triggering

5Set the time base to 200 µs/div, then center the signal on the displaywith the Delay knob (delay about 800 µs).6Press Main/Delayed , then press the D

Seite 70

8Press Main/Delayed , then press the Main softkey.9Use the horizontal vernier to change the time base to 1 µs/div, thencenter the signal on the displa

Seite 71 - Trigger types

Triggering on both TV Fields at the Same TimeThe HP 54645A/D Oscilloscopes can trigger on the vertical sync pulse in both TVfields at the same time. T

Seite 72

Measuring Waveform DataWith the HP 54645A/D Oscilloscopes, you can measure waveform datausing Voltage, Time, and Cursor keys. This section shows you h

Seite 73

Use the Entry knob to move the cursors. You can toggle the active cursorby pressing the cursor key while in the cursors menu.Cursor measurements are n

Seite 74

Cursors Track Delayed SweepExpand the display with delayed sweep, then characterize the event ofinterest with the cursors.Figure 5-18Making Measuremen

Seite 75 - To define an edge trigger

Moving Cursors TogetherPressing the time cursor softkey until both time cursors are highlightedallows you to move them together when rotating the Entr

Seite 76 - To define a pattern trigger

To make frequency measurements automaticallyThe automatic measurement capability of the HP 54645A/D Oscilloscopesallows you to easily make frequency m

Seite 77

•To find the Show Meas softkey, press the Next Menu softkey key.When the Show Meas softkey is turned on, cursors are displayed onthe waveform, and sho

Seite 78

To make time measurements automaticallyYou can measure the following time parameters with the oscilloscope:frequency, period, duty cycle, positive and

Seite 79

Items Supplied with the HP 54645A/D OscilloscopesFigure 1-3Getting StartedCheck package contents1-6

Seite 80

2Press Time .The first softkey menu appears with frequency, period, and duty cyclemeasurement choices.Source selects channel 1 or 2 for the time meas

Seite 81

4Press the Rise Time softkey.The oscilloscope automatically measures the rise time of the signal anddisplays the result on the display.Automatic Measu

Seite 82

To measure the duty cycle•Press Next Menu until you find the Duty Cy softkey, and press it.To measure the pulse widths1To measure the positive pulse w

Seite 83

To make voltage measurements automaticallyYou can measure the following voltage parameters automatically with theoscilloscope: peak-to-peak, average,

Seite 84 - Oscilloscope Operation

Voltage MeasurementsVoltage measurements do not work on digital channels on HP 54645D.1Connect a signal to the oscilloscope and obtain a stable displ

Seite 85 - Operation

Isolating Area for Vtop Voltage Measurement4Press the Next Menu softkey.A softkey menu appears with additional voltage measurement choices.Show Meas

Seite 86 - MegaZoom Concepts

6Using Option 005 EnhancedTV/Video Trigger (HP 54645A)

Seite 87 - Deep Memory

Using Option 005 EnhancedTV/Video TriggerYou can use the Option 005 Enhanced TV/Video trigger with theHP 54645A Oscilloscope. One of the first things

Seite 88 - Oscilloscope Responsiveness

Option 005 gives you an Enhanced TV/Video Trigger for theHP 54645A Oscilloscope, allowing highly detailed analysis ofTV waveforms. This option offers:

Seite 89 - Display Update Rate

To autoscale on a video signal1Use a cable to connect a TV signal to channel 1.2Press Slope/Glitch in the TRIGGER section of the front panel,and sele

Seite 90 - Display Modes

Optional Accessories for the HP 54645A/D OscilloscopesFigure 1-4Getting StartedCheck package contents1-7

Seite 91

To trigger on a specific line of videoTV triggering requires greater than 1/4 division of sync amplitude, eitherchannel 1 or channel 2 as the trigger

Seite 92

Triggering on Line 71Note the TV Grid displayed (selected from the Display menu).Table 6-1Line Numbers per Field for Each TV StandardTV Standard Field

Seite 93

To trigger on all TV line sync pulsesTo quickly find maximum video levels, you could trigger on all TV line syncpulses. When All Lines is selected as

Seite 94

To trigger on a specific field of the video signalTo examine the components of a video signal, trigger on either Field 1 orField 2. When a specific f

Seite 95

To trigger on all fields of the video signalTo quickly and easily view transitions between fields, or to find the amplitudedifferences between the fie

Seite 96 - Same Square Wave on HP 54645D

To trigger on odd or even fieldsTo check the envelope of your video signals, or to measure worst casedistortion, trigger on the odd or even fields. W

Seite 97 - HP 54645D Shows Glitches

If a more detailed analysis is required, then only one color field should beselected to be the trigger. You can do this by using the oscilloscope’s h

Seite 98

To analyze video waveforms with Option 005The combination of the TV trigger, delayed sweep, and automaticmeasurements allow this oscilloscope to preci

Seite 99

5Set the time base to 200 µs/div, then center the signal on the displaywith the delay knob (delay about 800 µs).6Press Main/Delayed , then press the D

Seite 101 - Pan and Zoom

Table 1-1Power CordsPlug Type Cable PartNo.Plug Description Lengthin/cmColorOpt 903 (U.S.A.)124V **8120-1378 Straight (NEMA5-15P*) 90/228 Jade GrayOpt

Seite 102

7Testing, Adjusting, andTroubleshooting

Seite 103

Testing, Adjusting, and TroubleshootingThis chapter explains the general procedures for solving problems,verifying correct oscilloscope operation, adj

Seite 104 - Run/Stop

Solving Problems with HP 54645A/DOscilloscopesThis chapter describes possible solutions if you encounter problemswhile making measurements with HP 546

Seite 105 - Main/Delayed

If there is no trace displayCheck that the power cord is connected to the oscilloscope and to a livepower source.Check that the front-panel power swit

Seite 106

If the trace display is unusual or unexpectedCheck that the Time/Div setting is correct for the expected frequencyrange of the input signals.The sampl

Seite 107

Check that the grabbers are securely connected to the nodes in thecircuit under test.Check that the circuit under test is powered on.You may have pres

Seite 108

Getting ServiceIf the oscilloscope is under warranty, you must return it toHewlett-Packard for all service work covered by the warranty. See "To

Seite 109

To return the oscilloscope to Hewlett-PackardBefore shipping the oscilloscope to Hewlett-Packard, contact your nearestHewlett-Packard Sales Office for

Seite 110

Verifying Oscilloscope PerformanceTo completely test and troubleshoot the oscilloscope, you will need tofabricate two test accessories:• The test conn

Seite 111

1Build the first test connector using a BNC connector and an 8-by-2section of Berg strip.aSolder a jumper wire to all pins on one side of the Berg str

Seite 112

User and Service GuidePublication Number 54645-97012December 1997 (pdf version Nov 1998)For Safety Information, Warranties, and Regulatory information

Seite 113 - Acquiring Data

Adjust the handle Connect the cables1Check to ensure that you received everything that is supplied withthe oscilloscope.2Connect the oscilloscope to p

Seite 114 - Memory Depth/Record Length

To test the digital channelsInstructions for HP 54645DThese instructions apply only to the HP 54645D Mixed-Signal Oscilloscope.Testing the acquisition

Seite 115

Testing Threshold AccuracyInstructions for HP 54645DThese instructions apply only to the HP 54645D Mixed-Signal Oscilloscope.Testing the threshold acc

Seite 116 - To take a single trace

To set up and connect the equipmentInstructions for HP 54645DThese instructions apply only to the HP 54645D Mixed-Signal Oscilloscope.1Turn on the tes

Seite 117 - To capture a single event

To verify threshold accuracyInstructions for HP 54645DThese instructions apply only to the HP 54645D Mixed-Signal Oscilloscope.1Press Label/Threshold

Seite 118 - Single

To check the output of the CALIBRATORIn this test, you measure the output of the rear-panel DC CALIBRATOROUTPUT with a multimeter. The CALIBRATOR is u

Seite 119 - To use Autostore

To verify voltage measurement accuracyIn this test, you verify the voltage measurement accuracy by measuring theoutput of a power supply using dual cu

Seite 120 - Autostore

1Set up the oscilloscope.aPress Setup , then press the Default Setup softkey.bAdjust the channel 1 Position knob to place the baseline atapproximately

Seite 121 - To erase the waveform display

7Continue checking the voltage measurement accuracy with theremaining power supply setting lines in the following table.Table 7-6 Voltage Measurement

Seite 122

To verify bandwidthIn this test you verify bandwidth by using a power meter and power sensor toset output of a signal generator at 1 MHz and at 100 MH

Seite 123

1Connect the equipment.aConnect the signal generator to the input of the power splitter.bConnect the power sensor to one output of the power splitter,

Seite 124

Power-on the oscilloscope1Connect the power cord to the rear of theHP 54645A/D, then to a suitable ac voltagesource.The HP 54645A/D power supplyautoma

Seite 125

8Change the time base to 5 ns/div.Wait a few seconds for the measurement to settle (the Av letters in the statusline indicate how much of the averagin

Seite 126

To verify horizontal ∆t and 1/∆t accuracyIn this test, you verify the horizontal ∆t and 1/∆t accuracy by measuring theoutput of a time mark generator

Seite 127

5Press Time , then press the Freq and Period softkeys.You should measure the following:Frequency 1 MHz, test limits are 995.9 kHz to 1.0042 MHz.Period

Seite 128

To verify trigger sensitivityIn this test, you verify the trigger sensitivity by applying 25 MHz to theoscilloscope. The amplitude of the signal is d

Seite 129

Internal TrigSensitivity1Press Setup , then press Default Setup.2Connect the signal generator to channel 1 using a 50-Ω feedthroughat the oscilloscope

Seite 130

External TrigSensitivity6Verify the external trigger sensitivity at 100 MHz, 100 mVp-p and at25 MHz, 50 mVp-p.Instructions for HP 54645A OnlyThis proc

Seite 131

To verify vertical output on Option 005This section applies only to the Option 005 Enhanced TV/Video Trigger.In this test, we will use the oscilloscop

Seite 132 - Table 4-1

HP 54645A/D Performance Test RecordSerial No. ______________________________________ Test by____________________________Test Interval ________________

Seite 133 - Table 4-2

Adjusting the OscilloscopeThis section explains how to adjust the oscilloscope so that it is atoptimum operating performance. You should perform the

Seite 134 - Label/Threshold

To adjust the power supplyThe power supply has a +5.1 V adjustment and a –5.25 V adjustment. Theother voltages are based on the +5.1 V adjustment. I

Seite 135

Using the digital probes to probe a circuit 1If you feel it’s necessary, turn off the powersupply to the circuit under test.Turning off power to the c

Seite 136 - Labels Off

2Measure the power supply voltages at E1 through E6 on the systemboard.The test points are not marked on the system board; see the following figurefor

Seite 137

If the +5.1 V measurement is out of tolerance, adjust the +5.1 V adjustmenton the power supply; if the –5.25 V measurement is out of tolerance, adjust

Seite 138

To perform the self-calibrationIn this procedure, you load the default calibration factors to give a knownstarting point for the firmware calibration.

Seite 139 - To define a new label

1Check the rear panel CALIBRATOR output level.If you are not sure how to check the CALIBRATOR, see "To check the outputof the CALIBRATOR."2L

Seite 140 - Press the

To adjust the low-frequency compensationIn this procedure, you adjust the low-frequency compensation adjustment foreach channel.Table 7-14Equipment Re

Seite 141 - To initialize the label list

6Adjust the channel 1 low-frequency compensation adjustment for asflat a pulse top as possible.Channel 1 is labeled "1" on HP 54645A and &qu

Seite 142 - HP 54645A/D Oscilloscopes

1Connect the pulse generator to channel 1, using a 50-Ω feedthroughat the oscilloscope input.Channel 1 is labeled "1" on HP 54645A and "

Seite 143 - Oscilloscopes

To adjust the displayThe display adjustments are optional and normally do not requireadjustment. You should use this procedure only for the few cases

Seite 144 - Capturing Data

8Adjust HB Cont (half-bright contrast) for the best contrast betweenthe half-bright and full-bright blocks.You can re-adjust Sub Bri, intensity contro

Seite 145 - To use delayed sweep

To adjust the Option 005 offset (R15) (HP 54645A)On the Option 005 board, there is only one adjustment. You must calibratethe oscilloscope before perf

Seite 146

6Connect the ground lead on each set of channels,using a probe grabber. The ground lead improvessignal fidelity to the instrument, ensuringaccurate me

Seite 147 - Asynchronous Noise in Pulse

Troubleshooting the OscilloscopeThe service policy for this instrument is replacement of defectiveassemblies. The following procedures can help isola

Seite 148

To construct your own dummy load1Obtain a connector that is compatible with the connector J3 on theLow Voltage Power Supply.2Connect the following loa

Seite 149

To check out the oscilloscope 1Is there an interface module connected to the rear panel of theoscilloscope?If yes, do the following steps. If not, go

Seite 150

6Disconnect the ribbon cable from the display board, then check thefollowing signals on the system board. Table 7-19 Display SignalsSignal Name Freque

Seite 151 - Display

9Is the fan running?If yes, go to "To run the internal self-tests." If not, do the steps below.The Low Voltage Power Supply has a thermal c

Seite 152

To clear error messagesOne or more of the following error messages may appear on the display(within the graticule space) as a result of internal NVRA

Seite 153 - To use channel math

3If any Fail message appears on the display after cycling power:aPress any menu key and cycle the power on the oscilloscope again.bHold the menu key d

Seite 154 - To use the Roll display mode

To check the Low Voltage Power Supply1Disconnect the power cord, remove the cover and set theoscilloscope on its side.CAUTIONIf you operate the oscill

Seite 155 - To use the XY display mode

If the +5.1 V measurement is out of tolerance, adjust the +5.1 V adjustmenton the power supply; if the –5.25 V measurement is out of tolerance, adjust

Seite 156 - Cursors

To run the internal self-tests1Perform the display test.aPress Print/Utility .bPress Service Menu, then Self Tst Menu and Display.cDo the half-bright

Seite 157

Using the oscilloscope probes1Connect the HP 10074A 1.5-meter, 10:1 oscilloscope probe BNCconnector to the channel 1 or 2 input on the HP 54645A or to

Seite 158 - Figure 5-13

4Perform the ROM testaPress the ROM softkey.bDoes the display message say "Scope self-test passed"?If yes, press any key to continue. If no

Seite 159 - To analyze video waveforms

To troubleshoot Option 005To isolate a malfunction to the Option 005 board, do the following:1Disconnect the three cables that connect the Option 005

Seite 160

8Replacement Parts

Seite 161

Replacement PartsThis chapter contains instructions for removing and orderingreplaceable assemblies. Also included is a parts list for the assembliesa

Seite 162

Removing and Replacing AssembliesRefer to the exploded view of the oscilloscope in this chapter fordetails on how the oscilloscope fits together. To i

Seite 163 - Measuring Waveform Data

To remove the cabinet1Turn off the oscilloscope and disconnect the power cable.2If a module is installed, remove it from the oscilloscope.3Using the T

Seite 164

To remove the front panel1Remove the intensity knob by pulling it straight out.2Disconnect the keyboard ribbon cable from the system board.3Remove the

Seite 165 - Cursors Track Delayed Sweep

To remove the display1Remove the front panel on the HP 54645A/D.2Disconnect the ribbon cable and the calibration cable from thedisplay.3Using the T10

Seite 166 - Moving Cursors Together

To remove the front-panel BNC connectorsThis procedure applies to channel 1 and 2 BNC connectors on HP 54645A,and to channel A1 and A2 BNC connectors

Seite 167

To remove the power supply1Remove the fan.2Disconnect the ground wire (green wire with the yellow stripe) fromthe deck.3Disconnect the ribbon cable fr

Seite 169

To remove the keyboard1Remove the front panel.2Remove all the knobs by pulling straight out.3Flex the bezel of the front panel to unsnap the small key

Seite 170

Replacing the Option 005 boardThe Option 005 board is only serviceable at the board level. No component level information is available. WARNINGHazardo

Seite 171

Replaceable PartsThe Hewlett-Packard exchange program allows you to exchange afaulty assembly with one that has been repaired and performanceverified

Seite 172

To order a replacement part•To order a part in the material list, quote the Hewlett-Packard partnumber, indicate the quantity desired, and address the

Seite 173

Exploded View of HP 54645A/D OscilloscopeFigure 8-3Replacement PartsTo order a replacement part8-13

Seite 174

Table 8-1 HP 54645A/D Oscilloscope Replaceable PartsReferenceDesignatorHP PartNumberQty DescriptionA1 0950-2735 1 Power supply assemblyA2 2090-0316 1

Seite 175 - softkey

ReferenceDesignatorHP PartNumberQty DescriptionMP17 54630-94605 1 Rear-panel labelMP18 (HP 54645A) 54645-42501 1 External trigger nutMP19 54630-61601

Seite 176 - TV/Video Trigger (HP 54645A)

Exploded View of Option 005 and Related Oscilloscope PartsFigure 8-4Replacement PartsTo order a replacement part8-16

Seite 177 - TV/Video Trigger

Table 8-2 Option 005 Replaceable PartsReference DesignatorHP PartNumberQty DescriptionA3 * 1 System Board (part of standard instrument)A5 54602-66502

Seite 179 - Slope/Glitch

2Front-Panel Overview

Seite 180 - TRIGGER

9Performance Characteristics

Seite 181

Performance CharacteristicsThis chapter lists the operating characteristics and performancespecifications for both the HP 54645A and HP 54645D Oscillo

Seite 182 - Triggering on All Lines

HP 54645A/D OscilloscopesNumber of Channels 2 Analog—labeled 1 and 2 on HP 54645A, A1 and A2 on HP 54645DProbes 10074ARange 1 mV/div - 5 V/divVernier

Seite 183 - Triggering on Field 1

HP 54645D Digital ChannelsNumber of Channels 16 Digital—labeled D0-D15Threshold Selections D0-D7, D8-D15Maximum Input Voltage ± 40 V peakThreshold Ran

Seite 184 - Triggering on All Fields

HP 54645A/D Oscilloscope Digitizing SystemVertical Resolution 8 bits on all settings except 7 bits on < 2 mV/div to1 mV/divHorizontal Resolution 50

Seite 185

HP 54645A/D Time BaseRange 2 ns/div to 50 s/div (HP 54645A)5 ns/div to 50 s/div (HP 54645D)Resolution 40 psVernier 1-2-5 increments when Off, 25 minor

Seite 186

HP 54645A/D Oscilloscope Trigger SystemSources Channels 1 and 2, line, and EXT (on HP 54645A only)Range (Internal) ± 6 divEdges Either rising or falli

Seite 187

HP 54645A Oscilloscope Trigger SystemTrigger System Source, Mode/Coupling, Slope/GlitchHoldoff, LevelChannels 1 and 2, Line, EXT (no trigger out)Eithe

Seite 188 - Second VITS Displayed

HP 54645A/D Setup FunctionsSave/Recall 10 front-panel setups can be stored and recalled fromnon-volatile memory.Trace Memory Two volatile pixel memori

Seite 189

HP 54645A/D General CharacteristicsEnvironmentalCharacteristicsMeets the requirements of MIL-T-28800D for Type III, Class 3, Style Dequipment as descr

Seite 190 - Troubleshooting

Front-Panel OverviewBefore you make measurements using the HP 54645A Oscilloscope andHP 54645D Mixed-Signal Oscilloscope, you must first set up the in

Seite 191 - •Adjusting the Oscilloscope

Option 005 General Performance CharacteristicsVideo Standards NTSC, PAL, PAL-M, SECAM, GenericVideo Trigger Modes:Line Field 1, Field 2, Alternate Fie

Seite 194 - If you can’t see a channel

MessagesThis chapter presents the various messages displayed by theinstrument, explains what they mean, and gives instructions forcorrective action if

Seite 195

MessagesAddress ErrorA processing fault occurred during operation. Cycle power to theoscilloscope, or perform a hard reset by cycling power with the t

Seite 196 - Getting Service

Delay time = When the Horizontal Mode is Delayed, this message appears while you areadjusting the Delay knob, to inform you of the current delay time

Seite 197

Measurement already selectedThis message appears when you select a single-channel measurement that isalready running. Note that measurement results fo

Seite 198

Privilege ViolationA processing fault occurred during operation. Cycle power to theoscilloscope, or perform a hard reset by cycling power with one of

Seite 199

The Label List has not been changedThis message appears when you press the No softkey during any step of labellist reinitialization. Initialization of

Seite 200 - To test the digital channels

Threshold is at limitThe threshold setting is used by the HP 54645D oscilloscope digital channelinput comparators to decide whether a signal is a logi

Seite 201 - Testing Threshold Accuracy

Viewing Signal Detail with Display ModeRemember how you had to constantly adjust the brightness on old analogscopes to see a desired level of detail i

Seite 202 - Threshold Test Setup

GlossaryThis glossary is organized into two parts: logic and scope terms.Logic terms apply only to theHP 54645D MSO (Mixed-SignalOscilloscope), and sc

Seite 203 - To verify threshold accuracy

Entered trigger operator isused in the advanced pattern trigger.“Entered” means that the triggerqualifier is satisfied as soon as the in-put waveforms

Seite 204

Sweep speed (time base) is theamount of time per division on thedisplay, and is shown on the statusline. When you vary the Time/Div set-ting while run

Seite 205

Autostore records the result ofevery acquisition on the display inhalf-bright tone. When Autostore ison, the oscilloscope updates pixel(display) memor

Seite 206

Delayed a horizontal mode thatgives an expanded view of the mainsweep.Delayed sweep is the oscilloscopedisplay function that magnifies thecontents of

Seite 207 - 1 mV/Div*

Glitch trigger mode is indicatedby a positive or negative pulse on thestatus line. In glitch capture mode,the oscilloscope uses memory re-sources to r

Seite 208 - To verify bandwidth

Noise Rej (noise reject) decreasesthe trigger sensitivity to reduce trig-gering on signal noise.Normal allows a waveform to bedisplayed if a trigger s

Seite 209

Run acquires data and displays themost recent trace.Run indicator is displayed at theright end of the status line when theacquisition is running.Sampl

Seite 210

Softkeys set up various optionsfor major functions. The softkeyschange depending on the functionyou select. Some softkeys have animmediate action, and

Seite 211

Trigger mode can be set to AutoLevel, Auto, or Normal. See “Normaltrigger mode” or “Auto trigger mode.”TV allows access to the TV modeand polarity key

Seite 212

The simplest averaging is “smoothing.” For example, the sample rate at aTime/Div setting of ≥ 2 µs/div allows the extra 5-ns samples to be smoothedtog

Seite 213 - To verify trigger sensitivity

IndexAaccessoriesoptional, 1–7supplied, 1–5 to 1–9accuracytesting thresholds, 7–12acquisitionand channels off, 2–24buffer, 2–28indicator for, 2–13 to

Seite 214

preset configuration, 4–48selecting, 2–26selecting for edge trigger, 3–16selecting for pattern trigger, 3–17test thresholds, 7–14testing acquisition s

Seite 215 - 25 MHz, 50 mV

EECL threshold, 4–51 to 4–52Edge, 3–16edge and pattern triggers combined, 3–22edge qualifier in pattern trigger, 3–17edge trigger, 3–16applications of

Seite 216

status, 2–13 to 2–14initializing label list, 4–58inputsetting signal thresholds, 4–51signals setting threshold, 4–51 to 4–52input channelsperformance

Seite 217

non-volatile settingslabel list, 4–56non-volatile setup memory, 4–44NTSC, 6–4, 6–6, 6–10 to 6–11Oodd field, 6–10 to 6–11On/Off"Off" channels

Seite 218 - Adjusting the Oscilloscope

sensing, 2–23tip design, 4–40probe cable, 1–11 to 1–12probe calibratorperformance characteristics, 9–9probe leadsremoving from cable, 1–12probesconnec

Seite 219 - To adjust the power supply

On/Off, 2–24Pattern, 3–17softwareBenchLink for Microsoft Windows, 1–7solutions to problemschannels, 7–5 to 7–6trace display, 7–4 to 7–5solving problem

Seite 220

trigger eventposition in acquisition memory, 4–22 to4–28trigger modeAuto, 2–13 to 2–14Normal, 2–13 to 2–14trigger not foundand auto trigger, 3–3trigge

Seite 221 - (top side of oscilloscope)

DECLARATION OF CONFORMITYaccording to ISO/IEC Guide 22 and EN 45014Manufacturer’s Name:Hewlett-Packard CompanyManufacturer’s Address:Colorado Springs

Seite 222

Product RegulationsSafetyIEC 1010-1:1990+A1 / EN 61010-1:1993UL 3111CSA-C22.2 No.1010.1:1993EMCThis Product meets the requirement of the European Comm

Seite 223

The HP 54645A/D at a GlanceDisplay shows current input signals•Eighteen (18) channels in main mode;2 scope and 8 timing channels indelayed mode (HP 54

Seite 224

Delayed SweepDelayed sweep on the HP 54645A/D is a simultaneous display of thewaveform at two different sweep speeds. Because of the deep memory in th

Seite 225

© Copyright Hewlett-Packard Company 1996-1997All Rights Reserved.Reproduction, adaptation, ortranslation without priorwritten permission isprohibited,

Seite 226 - Autoscale

Product WarrantyThis Hewlett-Packardproduct has a warrantyagainst defects in materialand workmanship for a periodof three years from date ofshipment.

Seite 227 - To adjust the display

HP 54645A/D Front PanelsHP 54645A Oscilloscope Front PanelTriggerKeysDisplay StorageKeysSoftkeysBrightnessControlPowerTriggerInputChannelInputs/Contro

Seite 228

HP 54645D Mixed-Signal Oscilloscope Front PanelTriggerKeysDisplay StorageKeysSoftkeysBrightnessControlPowerAnalog ChannelInputs/ControlsMeasurementKey

Seite 229 - (HP 54645A)

Description of Front-Panel AreasTable 2-1HP 54645A Front-Panel AreasFront-Panel Areas Functions Control KnobsDisplay Shows waveforms, measurement resu

Seite 230

Table 2-2HP 54645D Front-Panel AreasFront-Panel Areas Functions Control KnobsDisplay Shows waveforms, measurement results and instrumentconfiguration

Seite 231 - 19, and 20

Table 2-3HP 54645A/D Front-Panel SpecificsGeneral Front-Panel AreasDescription Applies to Unit:Waveform Display The waveform display area shows all ac

Seite 232

General Front-Panel AreasDescription Applies to Unit:Keypad and Softkeys Some keys, such as On/Off and Label also have special functionswhile in the m

Seite 233

General Front-Panel AreasDescription Applies to Unit:The Position knob moves the selected channel to a new verticalposition on the display. The HP 546

Seite 234 - Is the fan running?

Front-Panel Areas Description Applies to Unit:Status Indicators Status indicators include the following:Channels that are turned off are visible only

Seite 235 - To clear error messages

Front-Panel Areas Description Applies to Unit: Trigger condition shows the current trigger mode. The AutoLevel and Auto trigger modes display the &q

Seite 236

In This BookThis manual will guide you in using andservicing the HP 54645A Oscilloscopeand the HP 54645D Mixed-SignalOscilloscope. This manual is orga

Seite 237

Front-Panel Areas Description Applies to Unit:Channel Numbers The channel numbers are always shown along the left edge of thedisplay.HP 54645A/DChanne

Seite 238

Front-Panel OperationBy now you should be somewhat familiar with the VERTICAL orANALOG, STORAGE, HORIZONTAL, TRIGGER, and DIGITAL groups of front-pane

Seite 239 - Perform the keyboard test

To use oscilloscope channels to view a signal•To configure the instrument quickly, press Autoscale .•To undo the effects of autoscale, press Setup , t

Seite 240 - Perform the

To use digital channels to view a signal•To configure the instrument quickly, press Autoscale .•To undo the effects of autoscale, press Setup , then p

Seite 241 - To troubleshoot Option 005

To display signals automatically using Autoscale •To configure the instrument quickly, press Autoscale .Autoscale displays all connected signals that

Seite 242 - Replacement Parts

During Autoscale, the delay is set to 0.0 seconds, the Time/Div setting is afunction of the input signal (about 2 periods of the triggered signal on t

Seite 243

To adjust oscilloscope vertical scalingThis exercise guides you through the vertical keys, knobs, and status line.1Center the signal on the display us

Seite 244

•Toggle each softkey and notice which keys cause the status line to change.Channels 1 and 2 have a vernier softkey that allows the Volts/Divknob to ch

Seite 245 - To remove the cabinet

To Expand the Vertical SignalWhen changing the Volts/Div for analog channels, you can have the signalexpand (or compress) about the signal offset or t

Seite 246 - To remove the front panel

To turn analog channels on and off •To turn off analog channel #1, press 1 (on the HP 54645A),or A1 (on the HP 54645D), then press the leftmost sof

Seite 248 - Remove the system board

To turn digital channels on and off (HP 54645D)•To turn digital channels 0 through 7 on, press D0-D15 , thenpress the D0 - D7 On softkey. To turn th

Seite 249 - To remove the power supply

To rearrange the digital channels (HP 54645D) 1Turn the Select knob to choose the digital channel you want to move.Only channels that are currently on

Seite 250 - To remove the handle

To operate the time base controlsThe following exercise guides you through the time base keys, knobs, andstatus line.•Turn the Time/Div knob and notic

Seite 251

All events displayed left of the trigger point ▼ happened before the triggeroccurred, and these events are called pre-trigger information. You will f

Seite 252 - Replaceable Parts

To use the Entry and Select knobsThe Select knob (HP 54645D only) always changes the selected channel. Inmost menus, its order is screen-oriented. In

Seite 253 - To order a replacement part

To make cursor measurementsThe following steps guide you through the front-panel Cursors key. You canuse the cursors to make custom voltage or time m

Seite 254 - Figure 8-3

On HP 54645D, the Readout softkey gives a reading of the time orvoltage cursor measurement.•For the time cursor readout, you can select Time, Hex, Bin

Seite 255

To use delayed sweepDelayed sweep is a magnified portion of the main sweep. You can usedelayed sweep to locate and horizontally expand part of the ma

Seite 256

To modify the graticule1Press Display .2Press one of the Grid softkeys to define the graticule used for thewaveform area on the display.•Frame has a s

Seite 257 - Figure 8-4

To print the displayYou can print the waveform display and status line to an HP printer orEpson-compatible printer attached to one of these optional i

Seite 258

Contents1 Getting StartedPreparing the Oscilloscope 1–4Check package contents 1–5Adjust the handle 1–9Connect the cables 1–9Power-on the oscillos

Seite 259

3Triggering HP 54645A/D Oscilloscopes

Seite 260 - Performance Characteristics

Triggering HP 54645A/D OscilloscopesThe HP 54645A/D Oscilloscopes provide a full set of features to helpautomate your measurement tasks, including Meg

Seite 261

Selecting Trigger Modes and ConditionsThe trigger mode affects the way in which the oscilloscope searchesfor the trigger. Figure 3-1 shows the concept

Seite 262

When you select Single, the oscilloscope will fill acquisition memory, andcontinue flowing data through the pre-trigger buffer until the auto triggero

Seite 263 - HP 54645D Digital Channels

To use holdoffBy setting the Holdoff, you can synchronize triggers. The oscilloscope willtrigger on one edge of the waveform, and ignore further edges

Seite 264

To trigger on a complex waveformThe difficulty in viewing a complex waveform is triggering on the signal. Thesimplest trigger method is to trigger the

Seite 265 - HP 54645A/D Time Base

Triggering the HP 54645A OscilloscopeThe HP 54645A Oscilloscope allows you to synchronize the display tothe actions of the circuit under test by defin

Seite 266

To trigger on an edge1Press Source .2Select 1, 2, Ext, or Line as the trigger source.You can choose a channel that is turned off as the source for the

Seite 267

To use glitch triggering1Press Slope/Glitch . Then select Glitch.2Select Glitch Menu.3Define the polarity, duration qualifier, and duration time for

Seite 268 - HP 54645A/D Setup Functions

To use line triggering•Press Source . Then select Line.To use TV triggering1Press Source . Select 1 or 2 for the Trigger Source channel.2Press Slop

Seite 269

3 Triggering HP 54645A/D OscilloscopesSelecting Trigger Modes and Conditions 3–3To select a trigger mode: Auto Lvl, Auto, Normal 3–3To use holdoff

Seite 270 - Option 005 Trigger System

Triggering the HP 54645D Mixed-SignalOscilloscopeThe HP 54645D Mixed-Signal Oscilloscope allows you to synchronizethe oscilloscope display to the acti

Seite 271

Trigger typesThese trigger types are available:•Edge trigger•Pattern trigger•Advanced triggerChanges to the trigger specification are applied when you

Seite 272 - Messages

Glitch TriggerThe advanced glitch trigger menu lets you select the analog (A1 or A2) ordigital (D0-D15) channel on which to capture a glitch event. Yo

Seite 273

Specifying a Logic ThresholdYou can specify a threshold for the logic channels. See “To change the logicthreshold for input signals” in chapter 4 for

Seite 274

Sharing of SourcesThe source definitions for the simple pattern trigger are shared with thePattern 1 and Edge 1 sources of the advanced pattern trigge

Seite 275

To define an edge trigger1Press Edge .2Select a channel as the trigger source using either the Select knob orby pressing a Trigger Source softkey.You

Seite 276

To define a pattern trigger1Press Pattern .2For each analog channel (A1 or A2) or digital channel (D0-D15) youwant to include in the desired pattern,

Seite 277

To use glitch triggering1Press Advancd . Then select Glitch.2Select Glitch Menu.3Select the analog channel (A1 or A2) or digital channel (D0-D15)sour

Seite 278

To define an advanced pattern triggerThe advanced triggering operation depends on previous selections in theAdvanced Trigger menu.1Press Advancd . The

Seite 279

5Select the source(s) for the trigger operator by pressing the leftmostSource softkey until the desired source is highlighted.Refer to the table of tr

Seite 280

Run/Stop/Single/Autostore/Erase Operation 4–29Acquiring Data 4–30Memory Depth/Record Length 4–31To run and stop an acquisition 4–33To take a singl

Seite 281

9To set up a pattern trigger source, do the following: aSelect a channel for the source.•To set the condition for that channel, press the L, H, or X s

Seite 282

Setup rulesRemembering the following rules will make it easier to work with theadvanced triggering capabilities:•Duration operators are valid when onl

Seite 283

ExampleSuppose you have a pulse train where one of the pulses is of constantduration (4 µs) and all other signals of interest are repeated with that p

Seite 284

Source Operator andParametersPattern/Edge Pattern 1 Then L on channels D2, D1, and D0, others don’tcarePattern 2 L on channels 2 and 1, H on channel 0

Seite 285

4MegaZoom Concepts and Oscilloscope Operation

Seite 286

MegaZoom Concepts and OscilloscopeOperationThis chapter gives an overview of some oscilloscope concepts that willhelp you make better measurements. It

Seite 287

MegaZoom ConceptsMegaZoom technology combines the advantages of conventionaldeep-memory digital oscilloscopes with the responsiveness andupdate rate o

Seite 288

Deep MemoryThe primary advantage of a deep-memory oscilloscope is sustained samplerate, allowing you to capture at the maximum sample rate and capture

Seite 289

Oscilloscope ResponsivenessAn important element of an oscilloscope is how responsive it is to controlchanges. After a control is changed, such as the

Seite 290

Display Update RateIn analog oscilloscopes, the scope is “blind” while the electron beam is resetto the left side of the screen. Thus, the update rate

Seite 291

To use the XY display mode 5–14To analyze video waveforms 5–18Measuring Waveform Data 5–22To make cursor measurements 5–22To make frequency measur

Seite 292

Display ModesRemember how when using your analog oscilloscope you had to constantlyadjust the brightness? You did this to see the desired level of det

Seite 293

Normal ModeNormal mode is the display mode that you will use most of the time. For themajority of use models and signals, normal mode yields the best

Seite 294

Attempting to Capture Narrow PulseAs you change the Time/Div setting (zoom in), and look at a smaller timewindow around the trigger, the percentage of

Seite 295

Between the 500-µs/div picture (figure 4-1), and the 2-µs/div picture(figure 4-2), whether or not the pulse or some part of the pulse is displayed iss

Seite 296

Peak Detection Prevents AliasingAliasing produces misleading information. The peak detection circuitry in theHP 54645A/D helps to prevent aliasing by

Seite 297

Now consider the same waveform at 200 µs/div. All that remains is thesignal’s maximum and minimum values. This could lead you to believe thesignal tra

Seite 298 - DECLARATION OF CONFORMITY

Same Square Wave on HP 54645DThe maximum and minimum values are displayed in additional to somenormal samples. The additional normal samples here give

Seite 299 - Product Regulations

Peak Detect versus Normal on HP 54645D Digital ChannelsIn a digital system design, a glitch is an unintentional or unexpected signaltransition which m

Seite 300

Average ModeAveraging is a way to pull the signal out of the noise. Averaging works betterthan either a bandwidth limit or a brightness control. The a

Seite 301

Vectors On/OffOne of the most fundamental choices you must make about your display iswhether to draw vectors between the samples, or simply let the sa