Agilent 34410A/34411A 61/2 Digit High-Performance Multimeters Application NoteContentsIntroduction 1Features and Performance 2Applications and Soluti
The 34411A level trigger isspecified in units of any mea-surement function. In this case,a DCV measurement is beingmade at 50 k samples/s. With 1 mill
Borrowing from the earlier battery discharging example,an external trigger or level trigger (34411A only) can begin the process of makingany measureme
External DMM connected to switchThe fastest and most convenientway to scan through a list ofchannels is to hardware-couplethe DMM directly to a switch
Understanding Fast MeasurementsSome of the examples in the previous section requireprogramming the DMM to itsfastest measurement settingsand using the
How to set up the DMM for fast DC measurementsThere are a five key factorsthat control measurementspeed for DC measurements:• Autozero• Autorange• Ape
The aperture setting is set toMIN. For the 34410A, thiswould be 100 µs. For the34411A, 20 µs is the MIN. Thesame commands can be usedfor DCI, ohms, or
How to set up various triggering scenariosThe 34410A and 34411A have twotriggering models. The modelsare illustrated in Figure 10.When you use the SCP
System readings and throughput ratesThe 34410A and 34411A haveexcellent programming anddata retrieval performance.This performance is exhibitedin the
The Advantages of a Built-In Web ServerThe 34410A and 34411A have abuilt-in Web server that providesa very powerful configuration,diagnosis, and progr
ConclusionThe 34410A and 34411A DMMsoffer superior ease of use on thebench, and they offer blisteringperformance for test-systemapplications. Whether
The 34410A and 34411A arebackward compatible with theAgilent 34401A multimeter andsupport a 34401A emulationmode. You will find a detailedcomparison o
www.agilent.comFor more information on AgilentTechnologies’ products, applicationsor services, please contact your localAgilent office. The complete l
Key features for system useRemove the rubber bumpersand handle, and you have asystem-ready DMM that canoutperform even VXI and PXIinstrumentation. You
Applications and SolutionsWhether your applicationrequires general-purpose measurements, precision DC and AC measurements,waveform capture of mechanic
capacitance of the DMM andlead capacitance of the probescan be calibrated out using the built-in math null function,which subtracts the initial mea-su
Precision DCV measurementscombined with peak measurementsPower supplies often have ripple voltages that are ridingon top of the desired DC output.Thes
In manufacturing, the goal is to minimize test time, so thecombination of precision DCVand peak measurements cansignificantly reduce test timecompared
Testing fluorescent ballasts using direct-sampled ACVWhen a fluorescent lamp is off,the mercury/gas mixture withinthe tube is non-conductive. Whenpowe
Level triggering with scope-like waveform captureAerospace and automotiveapplications are replete withmechanical-electrical signals.There are mechanic