1 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service Manual850 FaultsFault 200—Pneumatics shutdown: faulty pneuma
850 FaultsFault 226—Front detector thermal shutdown10 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualF
11 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFaults 850Fault 230—Front inlet thermal shutdownThe f
850 FaultsFault 232—Back inlet thermal shutdown12 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFault
13 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFaults 850Fault 235—Aux 2 thermal shutdownFault 235—A
850 FaultsFault 238—Faulty line interrupt. Thermal shutdown.14 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Servic
15 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFaults 850Fault 242—Pneu aux module invalid constants
850 FaultsFault 247—Front inlet module: obsolete EEPROM16 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service Man
17 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFaults 850Fault 252—Front inlet: non-inlet moduleFaul
850 FaultsFault 257—Front inlet: invalid inlet module18 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service Manua
19 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFaults 850Fault 263—Host communications: GPIB (or HPI
850 FaultsFault 203—Signal DSP faulty2 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualThe sequence wou
850 FaultsFault 270—Front inlet: invalid module checksum20 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service Ma
21 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFaults 850Fault 279—Pneumatics aux: invalid constants
850 FaultsFault 288—B OIM not installed22 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service Manual
3 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFaults 850Fault 205—Signal DSP RAM brokenFault 205—Sig
850 FaultsFault 208—Signal path test failed4 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFault 208—
5 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFaults 850Fault 214—Front detector flame outIf the ele
850 FaultsFault 218—F TCD filament shorted6 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service Manualboard, or t
7 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFaults 850Fault 220—Heater over current. Thermal shutd
850 FaultsFault 221—Thermal shutdown (status message)8 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service Manual
9 of 22Jun 2001Error Codes and Internal DiagnosticsAgilent 6890 Gas Chromatograph Service ManualFaults 850Fault 223—Oven thermal shutdownIf the oven s