Agilent Technologies 8496G Wartungshandbuch

Stöbern Sie online oder laden Sie Wartungshandbuch nach Multimeter Agilent Technologies 8496G herunter. Agilent Technologies 8496G Service manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - 8494/95/96G/H Attenuators

Agilent Technologies8494/95/96G/H AttenuatorsOperating and Service ManualAgilent Part Number: 08495-90025Printed in USAPrint Date: August 2004Supersed

Seite 2 - ESTRICTED RIGHTS LEGEND

4 8494/95/96G/H Operating And Service ManualOverviewTable 3 8496G/H Attenuator Switching OrderEach solenoid requires a drive of 20 to 30 V with a swit

Seite 3

8494/95/96G/H Operating And Service Manual 5OverviewFigure 2 Typical Solenoid Coil Driver CircuitsCAUTION Do not exceed the RF power rating of 1 W av

Seite 4 - Warranty

6 8494/95/96G/H Operating And Service ManualSpecificationsSpecificationsFrequency Rangeand AttenuationAttenuation Accuracy (±dB): (Referenced from 0 d

Seite 5 - Service and Support

8494/95/96G/H Operating And Service Manual 7SpecificationsMaximum SWRMaximum ResidualAttenuationAttenuation Repeatability ±0.01 dB typical after 5 mil

Seite 6 - Before Applying

8 8494/95/96G/H Operating And Service ManualSpecificationsSolenoid Cable Connector Refer also to Figure 3.Figure 3 Solenoid Cable Pin ConfigurationSec

Seite 7

8494/95/96G/H Operating And Service Manual 9SpecificationsMinimum Life >5 million cycles per sectionSwitching speed Maximum 20 milliseconds includi

Seite 8

10 8494/95/96G/H Operating And Service ManualInstallationInstallationInitial Inspection Inspect the shipping container for damage. If the shipping con

Seite 9

8494/95/96G/H Operating And Service Manual 11InstallationInstallation InstructionsThe attenuators may be installed with or without the base. The base

Seite 10 - Overview

12 8494/95/96G/H Operating And Service ManualOperating InstructionsOperating InstructionsCAUTION Do not apply RF power greater than 1 W average, or 1

Seite 11

8494/95/96G/H Operating And Service Manual 13Operating InstructionsProcedure 1. Connect equipment as shown in Figure 4 on page 12 with the attenuator

Seite 12 - Specifications

ii 8494/95/96G/H Operating And Service ManualNoticeThe information contained in this document is subject to change without notice. Agilent Technologie

Seite 13 - (all models)

14 8494/95/96G/H Operating And Service ManualOperating InstructionsPerformance Tests The instrument can be tested to the accuracy of the “Specificatio

Seite 14 - Refer also to Figure 3

8494/95/96G/H Operating And Service Manual 15Replaceable PartsReplaceable PartsTable 5 lists the replaceable parts which are the only parts that can b

Seite 15 - Physical Characteristics

16 8494/95/96G/H Operating And Service ManualServiceServiceTroubleshooting Troubleshooting consists of performing the “Operator’s Check” on page 12. I

Seite 16 - Installation

8494/95/96G/H Operating And Service Manual 17Service4. Close the gap between coupling nut with the nut on the extended-sleeve connector.To disconnect

Seite 17 - Instructions

18 8494/95/96G/H Operating And Service ManualServiceCAUTION Do not use aromatic or chlorinated hydrocarbons, esters, ethers, terpenes, higher alcohol

Seite 18 - Operating Instructions

8494/95/96G/H Operating And Service Manual 19Service3. Allow the jaws of the tool used to close and pull straight back from the connector without twi

Seite 19

8494/95/96G/H Operating And Service Manual iiiWhat You’ll Find in This Manual…•Overview, page 1• Specifications, page 6• Installation, page 10• Operat

Seite 20

iv 8494/95/96G/H Operating And Service ManualWarrantyCustom systems are warranted by contractual agreement between Agilent Technologies and the custom

Seite 21 - Replaceable Parts

8494/95/96G/H Operating And Service Manual vService and SupportProduct maintenance agreements and other customer assistance agreements are available f

Seite 22 - Conductor Contact

vi 8494/95/96G/H Operating And Service ManualInstrument MarkingsSafety Earth GroundThis is a Safety Class I product (provided with a protective earthi

Seite 23 - To care for the device:

8494/95/96G/H Operating And Service Manual 1OverviewThis manual contains operating instructions for the Agilent 8494G/H, 8495G/H, and 8496G/H attenuat

Seite 24 - 7 mm Center Contact

2 8494/95/96G/H Operating And Service ManualOverviewFigure 1 Typical Four Section Attenuator Schematic DiagramTable 1 and Table 2 on page 3 and Table

Seite 25 - Connector Outer Shell

8494/95/96G/H Operating And Service Manual 3OverviewTable 2 8495G/H Attenuator Switching OrderTable 1 8494G/H Attenuator Switching Order8494G/HAttenua

Kommentare zu diesen Handbüchern

Keine Kommentare